Type tests performed on a personnel dosimetry system according to IEC 61066

Authors

  • Romel Castillo Instituto Peruano de Energía Nuclear (IPEN)
  • Paula Huamanlazo Instituto Peruano de Energía Nuclear (IPEN)
  • Enrique Rojas Instituto Peruano de Energía Nuclear (IPEN)

Keywords:

Personnel dosimetry, Calibration, Thermoluminescent dosemeters

Abstract

In this study, the verification of the Harshaw 6600 Plus TLD personal dosimetry system was made using the method of the IEC-61066 type tests and the recommendations of the ISO 4037 standards. For this purpose, five dosimeters were irradiated over a water phantom using an irradiator with a 137Cs source; five dosimeters as control were also used. The evaluated parameters were homogeneity, detection limit, residual reading, linearity, reproducibility, droppings and temperature and humidity variations. The obtained results show that the Harshaw 6600 TLD dosimetric system fulfills the IEC 61066 criteria.

Downloads

Download data is not yet available.

Published

2016-12-22

How to Cite

Castillo, R. ., Huamanlazo, P., & Rojas, E. (2016). Type tests performed on a personnel dosimetry system according to IEC 61066. Informe Científico Tecnológico, 15(2), 12–17. Retrieved from https://revistas.ipen.gob.pe/index.php/ict/article/view/9

Issue

Section

Nota técnica